Generic selectors
Exact matches only
Search in title
Search in content
Post Type Selectors
Search in posts
Search in pages
Filter by Categories
Author Reply
Book Review
Brief Communication
Case Report
Case Series
Clinical Case Report
Clinicopathological Conference
Commentary
Corrigendum
Editorial
Editorial – World Kidney Day 2016
Editorial Commentary
Erratum
Foreward
Guidelines
Image in Nephrology
Images in Nephrology
Letter to Editor
Letter to the Editor
Letters to Editor
Literature Review
Notice of Retraction
Obituary
Original Article
Perspective
Research Letter
Retraction Notice
Review
Review Article
Short Review
Special Article
Special Feature
Special Feature - World Kidney Day
Systematic Review
Technical Note
Varia
Generic selectors
Exact matches only
Search in title
Search in content
Post Type Selectors
Search in posts
Search in pages
Filter by Categories
Author Reply
Book Review
Brief Communication
Case Report
Case Series
Clinical Case Report
Clinicopathological Conference
Commentary
Corrigendum
Editorial
Editorial – World Kidney Day 2016
Editorial Commentary
Erratum
Foreward
Guidelines
Image in Nephrology
Images in Nephrology
Letter to Editor
Letter to the Editor
Letters to Editor
Literature Review
Notice of Retraction
Obituary
Original Article
Perspective
Research Letter
Retraction Notice
Review
Review Article
Short Review
Special Article
Special Feature
Special Feature - World Kidney Day
Systematic Review
Technical Note
Varia
View/Download PDF

Translate this page into:

Letters to Editor
30 (
1
); 55-57
doi:
10.4103/ijn.IJN_61_19

Discordance between Flow-Cytometry Crossmatch and Single Antigen Bead (SAB) Assay: An Uncommon Finding and its Resolution

Department of Transfusion Medicine, Medanta-The Medicity, Sector-38, Gurgaon, Haryana, India
Section Head, Flow Cytometry, Waikato Hospital, Hamilton, New Zealand
Address for correspondence: Dr. Aseem Kumar Tiwari, Department of Transfusion Medicine, Medanta-The Medicity, Sector-38, Gurugram - 122 001, Haryana, India. E-mail: aseemtiwari@gmail.com
Licence

This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.

Disclaimer:
This article was originally published by Wolters Kluwer - Medknow and was migrated to Scientific Scholar after the change of Publisher.

Sir,

An acute or hyper-acute rejection in any solid organ transplant is due to the presence of preformed anti-HLA antibodies.[1] These antibodies can be identified with cell-based assays; Complement-dependent Lympho-Cytotoxicity Crossmatch (CDCXM) and Flow Cytometry crossmatch (FCXM) and bead-based assays; Panel Reacting Antibodies (PRA) and Single Antigen Bead (SAB).[2] Using cell-based and bead-based assays in an algorithmic manner combines relative merits of each assay to our advantage and allows better interpretation of results. Routinely, in a pre-renal transplant work-up, commonest scenarios are where cell-based crossmatch and SAB are concordant, either negative or positive. We present a case that belongs to a third scenario, an unusual presentation, where CDCXM and FCXM were negative; SAB was positive and virtual crossmatch revealed Donor Specific Antibody (DSA).

A 25-year-old male patient suffering from end-stage renal disease was referred to our laboratory by the nephrologist for pre-transplant workup with his wife as prospective donor. As per the institutional protocol, low resolution HLA typing for class I (A and B) and class II (DR) antigens {polymerase chain reaction-sequence specific oligonucleotide probes (PCR-SSOP)} was performed for assessing relationship as a pre-requisite according to The Human Organ Transplant Act, India, 1994 and its amendments.[3] Anti-HLA antibody detection was negative by CDCXM and FCXM. CDCXM was performed using the standard National Institute of Health (NIH) technique and FCXM was performed for T-cell and B-cell on BD FACSVerse™ Flow cytometer (San Jose, CA, USA). Since the patient had history of blood transfusion, PRA was performed using Flow PRA Class I & II Screening Test kit (One Lambda, Inc., CA, USA) and it was found positive. As this scenario presented a discrepancy between cell-based and bead-based methods, repeat testing were done to rule out any technical errors. However, results remained the same. Decision was taken to perform SAB assay (LIFECODES LSA™ Kit, Immucor Transplant Diagnostics Inc., USA) on Luminex platform. SAB assay was positive for class I antibodies and negative for class II antibodies. Antibodies were identified against HLA-A*24:03 (MFI-11531) and HLA-A*24:02 (MFI-5252). Low resolution typing identified donor HLA-A allele as A*24 only. Therefore, high resolution typing for donor HLA-A locus was also done to identify complete antigen. High resolution typing revealed HLA-A*24:03 in donor and confirmed the presence of DSA.

To further understand and resolve this uncommon discrepancy between cell-based crossmatch and SAB, literature was reviewed.[4] Of all the possible mentioned reasons for such discordance, performing tests to negate pro-zone and post-zone phenomenon was undertaken and FCXM was repeated with dilutions of recipient's serum and donor's cells (dilutions 1:2 to 1:8). FCXM was found to be positive for T cells (median channel shift was 59; cut-off ≥26) and negative for B cells (median channel shift was 98; cut-off ≥110) with donor cells in 1:2 dilutions [Table 1 and Figure 1]. This positive result for T-cell FCXM corroborated with SAB results and resolved the discrepancy.

Table 1 Brief Case Illustration
HLA typing Results CDC FCXM SAB (donor specific antibody)
T-cell B-cell Class I Class II
Case Recipient A11,68 B52,52 DR04,14 Initial Negative Negative Negative AllelesHLA-A* 24:03 MFI11531 AllelesNo DSA MFINA
Age/Gender Sensitization
25/M Yes (blood transfusion)
Donor A03,24 B55,55 DR11,14 Serum dilutions Negative Negative Negative
Age/Gender Relationship
26/F Wife Cell dilutions Negative Positive Negative

CDC=Complement Dependent Cytotoxicity Crossmatch, FCXM=Flow Cytometry based Crossmatch, DSA=Donor Specific Antibodies

Figure 1
Flow Cytometry Crossmatch results at different serum and cell dilutions

All tests, CDCXM, FCXM, PRA, and SAB, are used to detect the presence of donor-specific anti-HLA antibodies (DSA) in the recipient and in most of the cases, results of all tests are in concordance. However, rarely there can be discordance. Literature review identified following reasons for discordance: Pro-zone/Post-zone effect, stability of antigens, antibody against denatured antigens v/s native antigen and allelic expression on the donor cell surface.

In the present case, post-zone phenomenon was responsible for the discordance between results of cell-based and bead-based assays. Excess of antigen inhibits lattice formation and subsequent agglutination between antigen-antibody may not occur, which can give false-negative results. Diluting the donor cells or increasing the serum-to-cell ratio can solve the problem.

The case also highlights the importance of performing cell-based and bead-based assay, in an algorithmic manner, in a pre-transplant work-up to rule-out any donor specific antibodies, especially if the recipient has history of any sensitizing event. It can be concluded that in these cases where cell-based assays results are discordant with the SAB assay results, post-zone effect should also be considered to confirm or rule out DSA.

Declaration of patient consent

The authors certify that they have obtained all appropriate patient consent forms. In the form the patient(s) has/have given his/her/their consent for his/her/their images and other clinical information to be reported in the journal. The patients understand that their names and initials will not be published and due efforts will be made to conceal their identity, but anonymity cannot be guaranteed.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

References

  1. , . Methodological aspects of anti-human leukocyte antigen antibody analysis in solid organ transplant. World J Transplant. 2014;4:153-67.
    [Google Scholar]
  2. , , . HLA antibody detection with solid phase assays: Great expectations or expectations too great.? Am J Transplant. 2014;14:1964-75.
    [Google Scholar]
  3. , , . “Transplantation of human organs and tissues Act-“Simplified”.”. Indian Journal of Transplantation 12.2 2018:84.
    [Google Scholar]
  4. , , , . Interpreting anti-HLA antibody testing data: A practical guide for physicians. Transplantation. 2016;100:1619-28.
    [Google Scholar]